ST108 Sample Collection Types and Procedure Overview

The sampling procedures below are provided as a summary for reference and are not exhaustive. You are advised to contact Legionella Control Systems for more comprehensive and detailed sampling procedures related to the AAMI ST108 standard.

Contact Legionella Control Systems with any questions about AAMI sample procedures, AAMI ST108 testing and AAMI ST108 risk assessments.

Below are groupings of ST108 test types, the collection device for each, and a simple description of the sample collection procedure. Also see AAMI ST108 Microbiological Quality Sampling Procedures for additional notes.

Heterotrophic Plate Count (HPC)

Collection Device: Sodium Thiosulfate Treated 120 ml BT

Summary of Procedure: HPC collected in 120 ml sealed and treated vessel. Return standard overnight on ice. Samples must be analyzed within 48 hrs. of collection.

Endotoxins

Collection Device: Sterile/Pyrogen Free Tube

Summary of Procedure: Endotoxin samples collected in 15 ml pyrogen free tube

pH, Conductivity, Chloride, Sulfate , Nitrate , Phosphate, Turbidity, Color

Collection Device: 125 ml plastic container

Summary of Procedure: Let water run for two minutes before collecting to ensure representative sample.

Total Alkalinity

Collection Device: 250 ml plastic container, no headspace

Summary of Procedure: Fill 250ml bottle completely to rim with no headspace to prevent CO2 interference.

Iron, Copper, Manganese, Aluminum, Zinc, Silicate, Total Hardness

Collection Device: 125 ml plastic container

Summary of Procedure: Let water run for one minute before collecting to ensure representative sample

Total Organic Carbon

Collection Device: 40 ml amber glass container

Summary of Procedure: Completely fill bottle. Ship in wrapping to prevent glass breakage.

Contact Legionella Control Systems at 888-416-8626 for more comprehensive and detailed sampling procedures related to the AAMI ST108 standard.