Testing Levels Needed to Meet ST108 Standard for Sterile Processing

When performing ST108 testing or ST108 risk assessments, below are the standard levels that need to be met for specific elements in the water. You can refer to these levels during your processing efforts for central sterile to set requirements and monitor progress. Also see our ST108 testing packages.

Water Quality Measurement Units Utility Water Critical Water Steam*
pH @ 25 °C: Units:pH Utility Water:6.5 – 9.5 Critical Water:5.0 – 7.5 Steam*”5.0 – 9.2**
Total Alkalinity Units:mg CaCO3/L Utility Water:<400 Critical Water:<8 Steam*”<8
Bacteria Units:CFU/mL Utility Water:<500*** Critical Water:<10 Steam*”N/A
Endotoxin Units:EU/mL Utility Water:N/A*** Critical Water:<10 Steam*”N/A
Total Organic Carbon (TOC) Units:mg/L (ppm) Utility Water:N/A Critical Water:<1.0 Steam*”N/A
Color and Turbidity Units:Visual Utility Water:Colorless, clear, without sediment Critical Water:Colorless, clear, without sediment Steam*”Colorless, clear, without sediment
Ionic Contaminants
Aluminum Units:mg/L Utility Water:<0.1 Critical Water:<0.1 Steam*”<0.1
Chloride Units:mg/L Utility Water:<250 Critical Water:<1 Steam*”<1
Conductivity Units:μS/cm Utility Water:<500 Critical Water:<10 Steam*”<10
Copper Units:mg/L Utility Water:<0.1 Critical Water:<0.1 Steam*”<0.1
Iron Units:mg/L Utility Water:<0.1 Critical Water:<0.1 Steam*”<0.1
Manganese Units:mg/L Utility Water:<0.1 Critical Water:<0.1 Steam*”<0.1
Nitrate Units:mg/L Utility Water:<10 Critical Water:<1 Steam*”<1
Phosphate Units:mg/L Utility Water:<5 Critical Water:<1 Steam*”<1
Sulfate Units:mg/L Utility Water:<150 Critical Water:<1 Steam*”<1
Silicate Units:mg/L Utility Water:<50 Critical Water:<1 Steam*”<1
Total Hardness Units:mg CaCO3/L Utility Water:<150**** Critical Water:<1 Steam*”<1
Zinc Units:mg/L Utility Water:<0.1 Critical Water:<0.1 Steam*”<0.1

* Steam parameters are for monitoring as steam condensate

** For local steam generation, the condensate pH should be 5.0 to 7.5. For boiler-treated steam, most boilers should be treated to maintain a condensate pH of 7.5 to 9.2

*** When Utility Water is used after chemical high-level disinfection as a final rinse, the bacteria should be <10 CFU/mL and endotoxin <10 EU/mL

**** If hardness is greater than 150 mg/L a water softener is recommended unless used for washing where the cleaning chemistry is capable of handling higher levels of hardness

Water quality management is crucial for every industry, including healthcare, hospitality, correctional facilities and manufacturing. Working with Legionella Control Systems gives you access to independent, expert insights and proven solutions for managing water safety. Contact Legionella Control Systems to schedule a site assessment today or call us for rapid testing.